Scientific Production Faculty

Determination of oxytetracycline in milk samples by polymer inclusion membrane separation coupled to high performance liquid chromatography



Rodríguez Ávila, José Antonio

2012

Pérez-Silva, I., Rodríguez J.A., Ramírez-Silva, M.T., Páez-Hernández, M.E. Determination of oxytetracycline in milk samples by polymer inclusion membrane separation coupled to high performance liquid chromatography, Analytica Chimica Acta, 2012, Vol. 718, p. 42-46, ISSN 0003-2670


Abstract


The determination of oxytetracycline in milk samples using a polymer inclusion membrane concept with high performance liquid chromatography (HPLC) was studied. The membranes developed are composed by cellulose acetate as polymer base, Cyanex 923 as carrier and o-nitrophenyl octyl ether as plasticizer. In the optimal conditions, the method exhibits good linearity in the range 0.030.20 mg L?1 with a limit of detection and quantification of 8.2 and 27.3 ?g L?1 respectively. The method was successfully applied to the analysis of milk samples with high selectivity.



Research Product




Related articles

Chemical studies of anthocyanins: A review

Synthesis and crystal structures of cis-palladium(II) and cis-platinum(II) complexes containing dipy...

Potentiometric Behavior of Diverse Polypyrrolesulphate Films Electro Synthesized on Graphite Epoxy ...

Sequential Injection Spectrophotometric Determination of Diclofenac in Urine and Pharmaceutical Form...

Potentiometric quantification of saccharin by using a selective membrane formed by pyrrole electropo...

Validation of a tubular bismuth film amperometric detector: Determination of diclofenac sodium by mu...

High-performance liquid magneto-chromatography

Synthesis, characterization, and crystal structures of n-alkyldiorganodithiophosphates RS2P(OC6H4)2

Potentiometric Behavior of Diverse Polypyrrole-sulphate Films Electro Synthesized on Graphite - Epox...

Development of a novel nitrate-selective composite sensor based on doped polypyrrole